LF-1.4 LF-1.6 LF-1.8
Balanced flux , Polyethesulfone membrane.
The dialyzer has the function of effectively eliminating carbamide,
creatinine and some medium molecular toxins,
and has excellent ultra-filtration function.
It applies to cure acute and chronic kidney failure patients caused
by severe injury, and poisoning.
Gama Sterilization
Balanced flux , Polyethesulfone membrane.
The dialyzer has the function of effectively eliminating carbamide,
creatinine and some medium molecular toxins,
and has excellent ultra-filtration function.
It applies to cure acute and chronic kidney failure patients caused
by severe injury, and poisoning.
Gama Sterilization
Type |
PES BF-1.4
|
PES BF-1.6
|
PES BF-1.8
|
Effective surface [sq.m]
|
1.4
|
1.6
|
1.8
|
Wall thickness
|
35μm
|
||
Internal diameter
|
200μm
|
||
Effective fiber length
|
240mm
|
||
Sterilization
|
Gamma
|
||
Max. TMP
|
500mm Hg
|
||
Dimensions
|
307mm×
38.5mm
|
307mm×
38.5mm
|
307mm×
42.5mm
|
Weight, average[g]
|
166
|
168
|
175
|
Performance Data Clearance
|
QB=200ml/min, QD=500ml/min,QF=10ml/min
|
||
Urea ml/min
|
184
|
190
|
194
|
Creatinine [ml/min]
|
172
|
178
|
182
|
Phosphate [ml/min]
|
150
|
158
|
167
|
Vitamin B12 [ml/min]
|
102
|
113
|
122
|
Ultrafiltration [ml/mmHg/h]*
|
18.6
|
20.1
|
22.5
|
Priming volume
blood compartment [ml]
|
82
|
92
|
110
|